Signal-space separation (SSS) and Maxwell filtering

This tutorial covers reducing environmental noise and compensating for head movement with SSS and Maxwell filtering.

As usual we’ll start by importing the modules we need, loading some example data, and cropping it to save on memory:

import os
import mne

sample_data_folder = mne.datasets.sample.data_path()
sample_data_raw_file = os.path.join(sample_data_folder, 'MEG', 'sample',
                                    'sample_audvis_raw.fif')
raw = mne.io.read_raw_fif(sample_data_raw_file, verbose=False)
raw.crop(tmax=60).load_data()

Background on SSS and Maxwell filtering

Signal-space separation (SSS) 1 2 is a technique based on the physics of electromagnetic fields. SSS separates the measured signal into components attributable to sources inside the measurement volume of the sensor array (the internal components), and components attributable to sources outside the measurement volume (the external components). The internal and external components are linearly independent, so it is possible to simply discard the external components to reduce environmental noise. Maxwell filtering is a related procedure that omits the higher-order components of the internal subspace, which are dominated by sensor noise. Typically, Maxwell filtering and SSS are performed together (in MNE-Python they are implemented together in a single function).

Like SSP, SSS is a form of projection. Whereas SSP empirically determines a noise subspace based on data (empty-room recordings, EOG or ECG activity, etc) and projects the measurements onto a subspace orthogonal to the noise, SSS mathematically constructs the external and internal subspaces from spherical harmonics and reconstructs the sensor signals using only the internal subspace (i.e., does an oblique projection).

Warning

Maxwell filtering was originally developed for Elekta Neuromag® systems, and should be considered experimental for non-Neuromag data. See the Notes section of the maxwell_filter() docstring for details.

The MNE-Python implementation of SSS / Maxwell filtering currently provides the following features:

  • Bad channel reconstruction

  • Cross-talk cancellation

  • Fine calibration correction

  • tSSS

  • Coordinate frame translation

  • Regularization of internal components using information theory

  • Raw movement compensation (using head positions estimated by MaxFilter)

  • cHPI subtraction (see mne.chpi.filter_chpi())

  • Handling of 3D (in addition to 1D) fine calibration files

  • Epoch-based movement compensation as described in 1 through mne.epochs.average_movements()

  • Experimental processing of data from (un-compensated) non-Elekta systems

Using SSS and Maxwell filtering in MNE-Python

For optimal use of SSS with data from Elekta Neuromag® systems, you should provide the path to the fine calibration file (which encodes site-specific information about sensor orientation and calibration) as well as a crosstalk compensation file (which reduces interference between Elekta’s co-located magnetometer and paired gradiometer sensor units).

fine_cal_file = os.path.join(sample_data_folder, 'SSS', 'sss_cal_mgh.dat')
crosstalk_file = os.path.join(sample_data_folder, 'SSS', 'ct_sparse_mgh.fif')

Before we perform SSS we’ll set a couple additional bad channels — MEG 2313 has some DC jumps and MEG 1032 has some large-ish low-frequency drifts. After that, performing SSS and Maxwell filtering is done with a single call to maxwell_filter(), with the crosstalk and fine calibration filenames provided (if available):

raw.info['bads'].extend(['MEG 1032', 'MEG 2313'])
raw_sss = mne.preprocessing.maxwell_filter(raw, cross_talk=crosstalk_file,
                                           calibration=fine_cal_file)

Out:

Maxwell filtering raw data
    Using loaded raw data
    Bad MEG channels being reconstructed: ['MEG 2443', 'MEG 1032', 'MEG 2313']
    Processing 204 gradiometers and 102 magnetometers
    Using fine calibration sss_cal_mgh.dat
        Adjusting non-orthogonal EX and EY
        Adjusted coil positions by (μ ± σ): 0.5° ± 0.4° (max: 2.1°)
    Automatic origin fit: head of radius 91.0 mm
    Using origin -4.1, 16.0, 51.7 mm in the head frame
        Using 87/95 harmonic components for    0.000  (72/80 in, 15/15 out)
    Processing 6 data chunks
[done]

Warning

Automatic bad channel detection is not currently implemented. It is critical to mark bad channels in raw.info['bads'] before calling maxwell_filter() in order to prevent bad channel noise from spreading.

To see the effect, we can plot the data before and after SSS / Maxwell filtering.

raw.pick(['meg']).plot(duration=2, butterfly=True)
raw_sss.pick(['meg']).plot(duration=2, butterfly=True)
  • ../../_images/sphx_glr_plot_60_maxwell_filtering_sss_001.png
  • ../../_images/sphx_glr_plot_60_maxwell_filtering_sss_002.png

Notice that channels marked as “bad” have been effectively repaired by SSS, eliminating the need to perform interpolation. The heartbeat artifact has also been substantially reduced.

The maxwell_filter() function has parameters int_order and ext_order for setting the order of the spherical harmonic expansion of the interior and exterior components; the default values are appropriate for most use cases. Additional parameters include coord_frame and origin for controlling the coordinate frame (“head” or “meg”) and the origin of the sphere; the defaults are appropriate for most studies that include digitization of the scalp surface / electrodes. See the documentation of maxwell_filter() for details.

Spatiotemporal SSS (tSSS)

An assumption of SSS is that the measurement volume (the spherical shell where the sensors are physically located) is free of electromagnetic sources. The thickness of this source-free measurement shell should be 4-8 cm for SSS to perform optimally. In practice, there may be sources falling within that measurement volume; these can often be mitigated by using Spatiotemporal Signal Space Separation (tSSS) 2. tSSS works by looking for temporal correlation between components of the internal and external subspaces, and projecting out any components that are common to the internal and external subspaces. The projection is done in an analogous way to SSP, except that the noise vector is computed across time points instead of across sensors.

To use tSSS in MNE-Python, pass a time (in seconds) to the parameter st_duration of maxwell_filter(). This will determine the “chunk duration” over which to compute the temporal projection. The chunk duration effectively acts as a high-pass filter with a cutoff frequency of \(\frac{1}{\mathtt{st\_duration}}~\mathrm{Hz}\); this effective high-pass has an important consequence:

  • In general, larger values of st_duration are better (provided that your computer has sufficient memory) because larger values of st_duration will have a smaller effect on the signal.

If the chunk duration does not evenly divide your data length, the final (shorter) chunk will be added to the prior chunk before filtering, leading to slightly different effective filtering for the combined chunk (the effective cutoff frequency differing at most by a factor of 2). If you need to ensure identical processing of all analyzed chunks, either:

  • choose a chunk duration that evenly divides your data length (only recommended if analyzing a single subject or run), or

  • include at least 2 * st_duration of post-experiment recording time at the end of the Raw object, so that the data you intend to further analyze is guaranteed not to be in the final or penultimate chunks.

Additional parameters affecting tSSS include st_correlation (to set the correlation value above which correlated internal and external components will be projected out) and st_only (to apply only the temporal projection without also performing SSS and Maxwell filtering). See the docstring of maxwell_filter() for details.

Movement compensation

If you have information about subject head position relative to the sensors (i.e., continuous head position indicator coils, or cHPI), SSS can take that into account when projecting sensor data onto the internal subspace. Head position data is loaded with the read_head_pos() function. The example data doesn’t include cHPI, so here we’ll load a .pos file used for testing, just to demonstrate:

head_pos_file = os.path.join(mne.datasets.testing.data_path(), 'SSS',
                             'test_move_anon_raw.pos')
head_pos = mne.chpi.read_head_pos(head_pos_file)
mne.viz.plot_head_positions(head_pos, mode='traces')
../../_images/sphx_glr_plot_60_maxwell_filtering_sss_003.png

The cHPI data file could also be passed as the head_pos parameter of maxwell_filter(). Not only would this account for movement within a given recording session, but also would effectively normalize head position across different measurement sessions and subjects. See here for an extended example of applying movement compensation during Maxwell filtering / SSS. Another option is to apply movement compensation when averaging epochs into an Evoked instance, using the mne.epochs.average_movements() function.

Each of these approaches requires time-varying estimates of head position, which is obtained from MaxFilter using the -headpos and -hp arguments (see the MaxFilter manual for details).

Caveats to using SSS / Maxwell filtering

  1. There are patents related to the Maxwell filtering algorithm, which may legally preclude using it in commercial applications. More details are provided in the documentation of maxwell_filter().

  2. SSS works best when both magnetometers and gradiometers are present, and is most effective when gradiometers are planar (due to the need for very accurate sensor geometry and fine calibration information). Thus its performance is dependent on the MEG system used to collect the data.

References

1(1,2)

Taulu S and Kajola M. (2005). Presentation of electromagnetic multichannel data: The signal space separation method. J Appl Phys 97, 124905 1-10. https://doi.org/10.1063/1.1935742

2(1,2)

Taulu S and Simola J. (2006). Spatiotemporal signal space separation method for rejecting nearby interference in MEG measurements. Phys Med Biol 51, 1759-1768. https://doi.org/10.1088/0031-9155/51/7/008

Total running time of the script: ( 0 minutes 6.244 seconds)